It is a kind of negative pressure device widely used in microbiology, biomedicine, gene recombinatio
Biological Safety Cabinet B2 series: 100% of the air is sent to operation area via downfall airflow and high efficient.
Complete system: control system composed of microcomputer chip
Composed of safety box, framework, fan, high efficient air filter and control system.
•Material: Safety box is made of high qualified cold-rolled plate and workroom is made of SUS 304 stainless steel.
•Air outlet is bilayer structure between workroom and outer shell.
•Air circulating system made of imported fan, high efficiency filter, and negative pressure system air passage. Inhaled air mixed with that in the workroom go through electrostatic box by driving of built-in centrifugal fan.
Model | BHC-1000 II B2 | BHC-1300 II B2 | BHC-1600 II B2 | |
Clean level | ISO 5( Class 100)/ISO 4(Class 10) | |||
Filter level/ Filter Efficiency | @0.3~0.12μm≥99.995%~99.999% | |||
Down flow Air Velocity | 0.25m/s~0.5m/s | |||
Inflow Air Velocity | ≥0.5m/s | |||
Noise | ≤67dB(A) | |||
Vibration Half Peak | ≤5μm | |||
Power | AC 220V/50Hz, single phase | |||
Biological protection air balance | Operator protection(Repeat for 3 times) | a. Total colony of impact mining pattern ≤10CFU./time b. Total colony of crack type sampling ≤5CFU./time | ||
Sample protection(Repeat for 3 times) | Total colony ≤5 CFU./time | |||
Cross-contamination protection(Repeat for 3 times) | Total colony ≤2 CFU./time | |||
Max. Consumption (KVA) | ≤1500KVA(with spare socket) | |||
Weight (Kg) | 150 | 200 | 250 | |
Work room size (W*D*H) (cm) | 100x57x63.5 | 130x57x63.5 | 160x57x63.5 | |
Outer size (cm) | 120x85.5x235 | 150x85.5x235 | 180x79.5x235 | |
Fluorescent tube Size &Quantity | 30W×② | 40W×② | 40W×② | |
UV Lamp Size & Quantity | 20W×① | 30W×① | 40W×① | |
Luminance | ≥650lx | |||
Diameter of air outlet (cm) | Φ20 | Φ25 | ||
Wind direction | Come in and out from the top |